![Beckman Coulter UniCel DxC Synchron 600 Instructions For Use Manual Download Page 380](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768380.webp)
Status-Cycle Count
Introduction
System Status and Commands
UniCel DxC Systems Instructions For Use A13914
Page 11-4
October 2005
Status-Cycle Count
Introduction
The Status-Cycle Count provides an approximation of an instrument's usage. This
information can be useful for estimating maintenance frequencies or anticipating
component failure. Cycle counts may be periodically collected by Beckman Coulter
using the instrument modem.
Displaying Cycle Count
Describing Cycle Count
Step
Action
1
Select the Status icon from the menu bar.
2
Select Count [F2].
3
A summary of the monitored areas appears.
Table 11.2
Counters
Counter
Description
Modular Chem
Each result for each cup is counted separately. Even if ORDAC
is triggered, the counter only increments for the one result.
ISE
A count of the number of samples run through the ISE module.
Whether it is a single chemistry per sample or a sample running
all five chemistries, the ISE cycle count will increment only
once per sample processed through the ISE module.
Cartridge Chem
Each cartridge chemistry reagent dispensed is counted.
Blade Count
• This is the number of caps pierced by the blade(s).
• For 1-Blade Thick CTS, the system sets this counter to zero
when the blade is replaced. To replace the blade for 1-Blade
Thick CTS, refer to
.
CTS Count
The total number of caps pierced by the Cap Piercer Module.
Sample Count
The total number of sample containers that have been scanned
by a sample ID bar code reader.