As-Needed/As-Required Maintenance
Replacing the CO
2
Membrane
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-84
October 2005
5
Inspect the membrane to verify that it is not broken, or wrinkled and is
centered properly with no uneven edges protruding. If membrane is not
centered correctly or is damaged, repeat Steps 1-4 with a new membrane.
Figure 10.60
6
CAUTION
Do not touch membrane surface when installing the quad-ring.
Reinstall the quad-ring (1) onto the membrane retainer. When installed, the
quad-ring must by on the very end of the CO
2
electrode. This ensures a
proper fluid seal when the electrode is installed in the flow cell.
Figure 10.61
Step
Action
, continued
(3 of 3)