![Beckman Coulter UniCel DxC Synchron 600 Instructions For Use Manual Download Page 327](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768327.webp)
As-Needed/As-Required Maintenance
Replacing the CO
2
Membrane
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-81
10
6
Remove electrode retainer nut (1). Remove membrane as follows:
• Separate the membrane retainer (3) assembly from the electrode.
• Remove the quad-ring (4) from the retainer. Inspect the electrode port if
the quad-ring is not on the electrode.
• Separate the membrane retainer from the membrane clamp (2) and pull
apart. Discard the used membrane.
Figure 10.55
Step
Action
, continued
(3 of 3)