Three-Month Maintenance
Replace Syringe Plungers
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-44
October 2005
Three-Month Maintenance
Replace Syringe Plungers
The one reagent and two sample syringe plunger rod assemblies should be replaced
every three months or when signs of wear (discoloration, flaking tips, etc.) are noticed.
The Syringe Plunger replacement procedures apply to the CC Reagent Syringe, the
MC Sample Syringe, and the CC Sample Syringe.
Figure 10.26 CC Reagent Syringe (500 µL)
CAUTION
To prevent damage due to electrical static discharge (ESD), wear a wrist
ground strap when performing this procedure. For further instructions, refer
to "
" in the beginning of this chapter.
A015942P.EPS