Monthly Maintenance
Replace Alkaline Buffer
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-21
10
Adjusting Fluid Level of the Alkaline Buffer Damper Assembly
Step
Action
1
Visually inspect the damper to verify that the fluid level rests between the
MIN and MAX lines, approximately one-third to one-half full.
If the fluid level is below the MIN line, proceed to Step 2.
If fluid level is above the MAX line, go to Step 8.
2
Remove the alkaline damper (1) from the holding clip and place the
damper in a horizontal position.
Note: The damper should be positioned so that the output line #29 (2),
leading from the damper to the flow cell, is oriented on top.
Figure 10.9
3
Prime alkaline buffer five times or as many times as necessary to fill the
damper one-half full. Access alkaline buffer priming by selecting:
• the Utils icon from the menu bar,
• <1> Prime,
• MC [F4],
• ISE CO
2
Alkaline Buffer.
Type 5 in the Number of primes to repeat field.
Select <Start Prime>. (Prime cycles are completed when Standby appears
in the system status.)
4
Observe the fluid level during prime cycles. As the damper becomes half
full, turn the damper to the upright position.
5
Wipe dry any reagent spill on the damper assembly.
(1 of 3)
A015938P.EPS