Weekly Maintenance
Check Reagent Levels
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-8
October 2005
Weekly Maintenance
Check Reagent Levels
Check the levels of these MC reagents: Wash Concentrate II, No Foam and CTS Auto-
Gloss (CTS option only). To check the levels, perform the following steps:
Step
Action
1
Select the Rgts/Cal icon from the menu bar.
2
Select the down arrow button to review the second page screen.
3
To access the modular reagents, open the left and middle doors.
Note: The CTS Auto-Gloss bottle is in a bracket on the inside of the middle
door.
4
Examine the reagent levels in the bottles and compare them with the
percentages on the screen (Wash, No Foam and Auto-Gloss).
5
If the volume shown does not match the actual reagent volume, enter the
correct volume for each reagent in the Vol field.
Note: There is a pop-up window when approximately 10% and 5% of an
MC reagent remains.
6
If necessary, load a new bottle of reagent.
Be careful when handling the bottle of CTS Auto-Gloss or No Foam.
Prevent spills. These solutions are extremely slippery and difficult to clean
from a floor.
Refer to CHAPTER 5, Reagent Load/Calibration,
this manual.
7
Close the left and middle doors.