User Defined Reagent Removal
Removing a User Defined Reagent
UniCel DxC Systems Instructions For Use A13914
User Defined Reagents
October 2005
Page 9-23
9
User Defined Reagent Removal
Removing a User Defined Reagent
Follow the steps below to explain how to clear a user-defined reagent.
A user defined reagent must be deconfigured from the chemistry menu prior to
removal.
Step
Action
1
Remove UDR reagent cartridge from system.
2
Delete UDR information from the following areas: reference ranges,
calibration assignments, sample programming and quality control.
3
Select the Setup icon from the menu bar.
4
Select <2> Chem Config/Sample Type.
5
Select the Option No. of the chemistry to be deleted.
6
Select [F6] Delete to remove the UDR from the chemistry menu.
7
Select the Setup icon from the menu bar.
8
Select <15> User-Defined Chemistries.
9
Select the Option No. of the chemistry to be deleted.
10
Select Delete [F2].