Within-Lot Calibration
Within-Lot Designations
UniCel DxC Systems Instructions For Use A13914
Reagent Load/Calibration
October 2005
Page 5-23
5
Figure 5.11
Within-Lot Calibration Status Screen
Within-Lot Designations
The designations wf, w, and ws follow the reagent lot numbers of those chemistries
selected for within-lot calibration. These designations are defined below.
E015927S.EPS
Table 5.4
Definitions of Within-Lot Designations
Designation
Description
wf
Reagent has been enabled for within-lot calibration and the
cartridge has been "freshly" loaded. Any wf-designated cartridge
can be used to establish a within-lot calibration factor. A
cartridge receives this designation if:
• it was loaded onto the system for the first time after within-lot
calibration was turned on.
• no more than 8 hours have lapsed since it was first loaded onto
the system.
w
Reagent has been enabled for within-lot calibration. A cartridge
receives this designation if:
• it was loaded for the first time after a within-lot calibration was
established.
• more than 8 hours have lapsed since it was first loaded onto the
system.
(1 of 2)