![Beckman Coulter UniCel DxC Synchron 600 Instructions For Use Manual Download Page 119](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768119.webp)
Special Calculations Definition
Beckman Coulter Predefined Special Calculation Formulas
UniCel DxC Systems Instructions For Use A13914
System Setup Options
October 2005
Page 4-27
4
Beckman Coulter Predefined Special Calculation Formulas
Crea Clearance (1)
mL/min
X.X
T Urine
Crea Clearance (2)
mL/sec
X.X
T Urine
ApoA/ApoB Ratio
None
X.X
Serum
ApoB/ApoA Ratio
None
X.X
Serum
Hemoglobin A1c
%
X.X
Serum
Table 4.2
Beckman Coulter Predefined Special Calculations Defaults
, continued
Calculation
Unit
Precision
Sample Type
(2 of 2)
Table 4.3
UniCel DxC 600 Predefined Special Calculation Formulas
Calculation
Formula
Osmality (1)
(1.86 × NA) + (GLUCm/18) + (BUN/2.8) + 9
Osmality (2)
(1.86 × NA) + (GLUCm/18) + (UREA) + 9
Anion Gap (1)
NA - (CL + CO
2
)
Anion Gap (2)
(NA + K) - (CL + CO
2
)
A/G Ratio
ALB / (TP - ALB)
Indirect Bili
TBIL - DBIL
BUN/Crea Ratio
BUN/CREA
Urea/Crea Ratio
UREA/(CREA × 0.0884)
Free Thyroxine
T4 × (TU/0.342)
Crea Clearance (1)
[(U × V)/P] × (1.73/A) reported in mL/min
Crea Clearance (2)
[(U × V)/P] × (1.73/A) reported in mL/sec
ApoA/ApoB
ApoA/ApoB
ApoB/ApoA
ApoB/ApoA
HbA1c
A1c/Hb × 100