Reference/Critical Ranges Setup
Reference Range and Critical Ranges
System Setup Options
UniCel DxC Systems Instructions For Use A13914
Page 4-24
October 2005
Reference/Critical Ranges Setup
Reference Range and Critical Ranges
Reference and Critical Range Setup options allow the operator to define the normal
and critical ranges for each analyte by age group, gender and sample type. Up to 32
age ranges may be defined. In addition, the operator may select one reference range as
the default.
Each facility should establish reference ranges based on its own population
demographics.
The Critical Range limits are used in conjunction with the Critical Results Rerun
option. When the limits of the critical range defined by the operator are exceeded, the
system will automatically rerun the analyte. The operator may delete one result or
send both to the host.
The Critical Result Rerun option should not be enabled for analytes with POS or NEG
result reports such as the drug of abuse tests (DAT).
From the Setup screen, select <10> Reference/Critical Ranges to define, edit, delete,
view or set a default.
Reference and Critical Ranges may be viewed while the system is running
The system must be in Standby or Stopped to edit these ranges or to enable/disable
Critical Results Rerun.