System Description
Cartridge Chemistry: Calibration Theory
1
Table 1.18
Math Models for Non-Linear Chemistries
Type
Formula
Model #1
Math Model #1 is the four-parameter log-logit function most commonly used with
reagents that use antibodies.
Sample values are determined using the calculated curve parameters and the math
model. Values may be calculated directly as this model can be solved for concentration.
Model #2
Math Model #2 is a five-parameter logit function.
This function cannot be solved directly for concentration. The instrument uses an
iterative method to determine the sample value.
Model #3
Math Model #3 is a five-parameter exponential function.
This function cannot be solved directly for concentration. The instrument uses an
iterative method to determine the sample value.
Models #4
through #7
These Models are reserved for future development.
Model #8
Math Model #8 is an alternative to model #2, the five-parameter logit function.
This function cannot be solved directly for concentration. The instrument uses an
iterative method to determine the sample value.
E014421L.EPS
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......