A13914AF
11-5
Utilities
Prime
11
Priming Hydropneumatic Subsystems
The Hydropneumatics Subsystem allows independent priming of each of its components. Follow
the steps below to prime any combination of the components.
1
Select:
•
Utils
from the menu bar,
•
1 Prime
,
OR
Type
1
in the
Option Number
field.
•
Hydro F3
.
2
Select any or all of the hydropneumatic subsystems.
3
Select
Start Prime
. Although the instrument status remains in
Standby
, the hydro prime will
begin.
Priming MC Subsystems
MC Subsystem Prime allows the priming of the cup modules, the ISE flow cell and EIC, and the
Sample Delivery Subsystem.
•
If a cup selection is made and neither reagent or DI water is specified, only DI water will be
primed into the cup.
•
If both reagent and DI water are selected, both will be primed and the cup(s) will be left full of
DI water.
•
Whenever Acid is primed, Reference must also be primed to prevent acid shocking the
electrodes. Therefore there is no Acid-only option.
•
Only one ISE selection can be made at a time. However the
ISE All
option can be used to select
the entire list.
1
Select:
•
Utils
from the menu bar,
•
1 Prime
,
OR
Type
1
in the
Option Number
field.
•
MC F4
.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......