A13914AF
10-4
System Status and Commands
Power Subsystems
2
Select the
Temp
tab. A summary of the monitored areas appears.
Power Subsystems
Introduction
Status for the Power Subsystem shows:
•
the actual voltage for each power distribution bus with the valid range.
•
the status of the power supplies, peltiers and fans.
•
an indication of when a parameter is out of limits.
A
yellow
highlight indicates a warning condition. A
red
highlight indicates an error condition.
Also shown is the status of the eight power supplies - OK, Warning, or Error with the same highlight
colors.
Show the Power Subsystem Status
1
Select
Status
from the menu bar.
2
Select the
Power
tab. A summary of the monitored areas appears.
Hydropneumatics Subsystem
Introduction
Status for the Hydropneumatics Subsystem shows the status for all of the reservoirs, canisters, and
sumps. A
yellow
highlight indicates a warning condition and a
red
highlight indicates an error
condition.
The following parameters are shown:
•
external DI water system status, if enabled.
•
status of the various hydro float switches.
•
DI water resistivity measured on the instrument and its valid range.
•
actual air and vacuum pressures and their valid ranges.
The same color highlights apply.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......