Maintenance
As-Needed/As-Required Maintenance
CAUTION
Pinch hazard. Keep fingers clear of the ISE module as you lower it.
27
Lower the ISE module by releasing the locking pin. Be careful NOT to pinch tubing as the module
is lowered.
28
Reinstall the ISE module cover.
29
Calibrate the ISE chemistries.
Clean the CO
2
Alkaline Buffer Lines
Bleaching of the Alkaline Buffer reagent straw and line is recommended to clean the reagent line
whenever the buffer is prematurely decolorizing (fading).
Prepare a 10% bleach
*
solution (one part 5.25% sodium hypochlorite (PN A32319), combined with
nine parts deionized water). Use within 24 hours.
CAUTION
To prevent damage due to electrical static discharge (ESD), wear the wrist ground
strap when you perform this procedure. For further instructions, refer to
in the beginning of this chapter.
1
Loosen the two Phillips screws from the ISE module cover and remove the cover.
2
Release the locking pin on the ISE module and lift the module until it clicks into a raised, locked
position.
3
Disconnect line #33 from the alkaline buffer reagent bottle and place the line in a small
container. (The container is used to catch the line drainage while priming).
4
Unscrew and remove the alkaline buffer reagent bottle cap and insert the straw into a container
of 10% bleach solution.
*
DO NOT use bleach that contains additives (for example, Ultra Bleach Advantage).
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......