Maintenance
As-Needed/As-Required Maintenance
11
Calibrate all the ISE chemistries.
Clean the EIC Port
Flushing of the EIC ports should be performed if the Electrolyte Injection Cup (EIC) becomes
plugged and overflows, or as part of troubleshooting for erratic electrolyte results. The following
items should be available before starting this procedure:
•
10% bleach
*
solution (one part 5.25% sodium hypochlorite (PN A32319), combined with nine
parts deionized water). Prepare fresh. Use within 24 hours.
•
70% isopropyl alcohol
•
20 or 30 mL irrigation syringe
CAUTION
To prevent damage due to electrical static discharge (ESD), wear the wrist ground
strap when you perform this procedure. For further instructions, refer to
in the beginning of this chapter.
1
Loosen the two Phillips screws on the ISE module cover and remove the cover.
2
Release the locking pin on the ISE module and lift the module until it clicks into a raised, locked
position.
3
Prepare ISE for maintenance and disable autoprime.
•
Select:
—
Utils
from the menu bar,
—
2 Maintenance
,
—
3 ISE Service
.
•
Select
Continue
to prepare for maintenance.
*
DO NOT use bleach that contains additives (for example, Ultra Bleach Advantage).
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......