A13914AF
9-61
Maintenance
Six-Month Maintenance
9
2
Press the latch (1), which is located on the connector, and pull out to disconnect sensor pin lead
from connector panel.
Figure 9.41
3
Unscrew the sensor retainer nut (1) and withdraw the sensor assembly from chemistry reaction
cup.
Verify that the gasket (2) from the removed sensor has not been left in port opening.
Figure 9.42
4
Separate the sensor from retainer nut. Discard the old sensor and gasket.