Maintenance
Three-Month Maintenance
Figure 9.28
MC and CC Sample Syringes (100 μL)
Prepare for Maintenance Mode and Remove the Plunger
Before removing the plunger, the plunger rods must be fully extended to the bottom of the syringe
barrels. Follow the steps below to activate the automated Syringe Rod Replacement procedure.
IMPORTANT
The system status must be in
Standby
to access
Maintenance
mode.
1
Select:
•
Utils
from the menu bar,
•
2 Maintenance
,
•
7 Syringe Rod Replacement
.
2
Select
Continue
to prepare the syringe rod assemblies for maintenance.
OR
Select
Cancel
to immediately return to the Maintenance menu without performing syringe rod
maintenance.
1.
MC Sample Syringe
2.
CC Sample Syringe
3.
Syringe Attached to Luer-lock Fitting
A011873P.EPS
1
2
3