Maintenance
Weekly Maintenance
9
Calibrate Total Protein and Albumin chemistries and run QC materials.
IMPORTANT
A lamp calibration should be performed after Total Protein and Albumin cup maintenance
on a
MONTHLY
basis. You are not required to perform the ALBm and TPm lamp calibration following
the
Weekly
maintenance.
Clean Probes (Exterior)
Follow the steps below to clean the exterior of the sample and reagent probes. Periodic wiping of
the probes is needed to clean the entire (upper and lower) probe area.
CAUTION
To prevent damage due to electrical static discharge (ESD), wear the wrist ground
strap when you perform this procedure. For further instructions, refer to
in the beginning of this chapter.
1
Press the
STOP
button on the instrument to place the system into
Stopped
state.
The probes are ready for manual maintenance when system status is
Stopped
.
2
Wipe the outside and bottom of any exposed portions of the sample and reagent probes, and all
of the cuvette wash station probes. Use a lintless tissue moistened with 70% isopropyl alcohol
or a 70% isopropyl alcohol pad.
NOTE
Use a separate pad for each probe.
Do not pull or exert pressure on probes, which can cause bending.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......