A13914AF
8-13
User Defined Reagents
Error Detection Limits
8
Usable Result Range
Usable result range specifies the analytical range of the reagent. Results are suppressed and flagged
as out-of-instrument range low (OIR-LO) or out-of-instrument range high (OIR-HI), respectively, if
these ranges are exceeded.
If any parameter is edited, the instrument automatically changes the Reportable Range information
from System Setup to the Usable Range defined for the UDR. The Reportable Range must be re-
entered if the values are different from the Usable Range.
Error Detection Limits
Reagent Blank (Absorbance or Rate)
Reagent blank specifies the minimum and maximum absorbance or rate (method dependent) for
the reagent blank measured during the blank read window. If these values are exceeded, the results
are suppressed and flagged as blank absorbance/rate high (BL ABS HI or BL RATE HI) or blank
absorbance/rate low (BL ABS LO or BL RATE LO).
Mean deviation is a measure of noise in the read window.
Reaction (Absorbance or Rate)
Reaction specifies the minimum and maximum absorbance or rate (method dependent) measured
during the reaction read window.
•
For upgoing rate or endpoint reactions, if the reaction absorbance is less than the Low Limit
then results are suppressed and flagged as RX ABS LO. Likewise, if the High Limit is exceeded,
the result is suppressed and flagged RX ABS HI.
•
For downgoing rate methods, the Low Limit indicates the maximum rate observed during the
reaction read window.
•
For downgoing endpoint methods, the Low Limit indicates the maximum absorbance observed
during the reaction read window.
Table 8.26
Usable Result Range
Allowable Entry
Lower Limit
0.000 to 99999.999
Upper Limit
0.000 to 99999.999
Table 8.27
Allowable Entries for Reagent Blank
Low Limit
-1.500 to 2.200
High Limit
-1.500 to 2.200
Mean Deviation
0.000 to 2.200
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......