A13914AF
8-1
CHAPTER 8
User Defined Reagents
Overview
The user may define up to 100 chemistries on the UniCel DxC 600/800 using an endpoint or rate, and
calibrated or non-calibrated methodology. A set of parameters must be defined which fully
characterizes an analyte. When the parameters are defined, they are stored in memory according
to the test name designated in System Setup. Chemistries can then be configured on the test
selection menu for programming along with the Beckman Coulter-defined chemistries. When
configured, quality control, reference ranges, and special calculations may be used with these
configured chemistries.
In addition, up to 10 of the 100 user defined chemistries may be configured with the Expanded User
Defined Chemistry feature. Refer to the end of this chapter for further information.
NOTE
The sample type for UDR must be serum, plasma, CSF or urine. Do not use any other sample type.
Requirements and Precautions
Minimum Requirements for Operation
The following parameters must be defined to use the User-Defined Reagent feature.
•
chem name
•
primary wavelength
•
secondary wavelength (except for the LPIA Module)
•
reagent dispense volume
•
sample volume
•
blank start and blank end read times
•
reaction start and reaction end read times.
If additional parameters are defined, Exit Check criteria must be passed. Refer to
in this chapter.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......