Reagent Load/Calibration
Reprint Calibration Reports
4
4
Move cursor to the slope and/or offset to be modified.
OR
Press
(
Page Up
)
or
(
Page Down
)
from the keyboard or use the screen scroll bar to locate the
Chem. Type the desired value in the appropriate
Slope
and
Offset
fields.
•
If a previously modified slope or offset is to be returned to the original default value for that
calibrator or chemistry, select desired chemistry, move cursor to the applicable field and
select
Restore Defaults
.
5
Select
Close
to exit screen.
Reprint Calibration Reports
Introduction
The five most recent calibration reports for a given calibrator (chem position) may be printed for
an historical perspective.
Reprint Current Calibration Reports
1
Select
Rgts/Cal
from the menu bar.
2
Select the chem position(s) to print.
NOTE
When requesting reports of recent MC calibrations, select only one chemistry per report format.
This prevents duplicate copies. For example, if NA is selected, the report contains all five ISE
chemistries.
3
Select
Options F6
.
4
Select
Print Cal Report
from the pull-down menu.
5
Select
Current Cal Report
from the
Print cal report
dialog box. The report(s) automatically prints.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......