A13914AF
4-28
Reagent Load/Calibration
Calibrator Set Point Modifications
Calibrator Set Point Modifications
Introduction
The Calibrator Set Point Modifications option allows editing of the programmed calibrator set
points. This option can be used to alter any chemistry provided it:
•
has calibration data.
•
uses three or less calibrator levels.
•
is configured on the system.
The user-defined chemistry set points may be modified within the user-defined parameters setup.
Modify Calibrator Set Points
IMPORTANT
Altering a programmed calibrator set point is not usually justified. However, there may be a
situation when a calibration value needs to be reassigned. Set Point Modification should not be used
routinely to correct instrument correlation. The magnitude of error which is deemed acceptable when
modifying a set point is totally under the control of the laboratory and, therefore, the Set Point
Modification function should be used with caution.
1
Select
Rgts/Cal
from the menu bar.
2
Select
Options F6
.
3
Select
Modify set points
from the pull-down menu.
4
Select
Next
or
Prev
to locate the calibrator of interest.
OR
Select the
Calibrator Name
pull-down menu at the top of the screen to view the List of
Calibrators.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......