A13914AF
4-27
Reagent Load/Calibration
Calibration Acceptance Limits
4
Calibration Acceptance Limits
Introduction
Calibration absorbance or ADC values are compared to preprogrammed back-to-back, span, and
range limits to determine acceptable performance. If calibration errors in accuracy, precision,
sensitivity, or linearity are detected by the system computer, error flags are generated. Calibration
Acceptance provides a reference to the acceptable limits for calibration.
View Calibrator Acceptance Limits
1
Select
Rgts/Cal
from the menu bar.
2
Select
Options F6
.
3
Select
Calibrator acceptance limits
from the pull-down menu.
Chemistries appear with chemistry name, setpoint value(s), and appropriate limits.
4
Select
Next
or
Prev
to locate the calibrator of interest.
OR
Select the
Calibrator Name
pull-down menu at the top of the screen to view the List of
Calibrators.
5
Select
Close
to exit the screen.