A13914AF
4-15
Reagent Load/Calibration
Reagent and Calibration Status Warnings
4
Reagent and Calibration Status Warnings
Introduction
When a reagent or chemistry calibration needs attention, a yellow or red warning highlight appears
in the applicable place on the
Reagent Status/Calibration Status
dialog box.
The highlight color depends on the reagent or calibration status.
Conditions that Cause a Calibration Status Warning
The table below lists the Calibration Status conditions and their related highlights.
Cal Overridden
Operator-initiated; results are based on the failed calibration.
Chem Bypassed
Operator-initiated; allows analysis to proceed on all other chemistries without
having to deprogram the bypassed chemistry. A bypassed chemistry can be
recalibrated, but not extended or overridden.
Cal Failed
Indicates that calibrator values for a chemistry have exceeded calibration
acceptance limits.
Cal Timed Out
Indicates time remaining has exceeded the recommended calibration period.
Cal Time Extended
Operator-initiated; indicates that the recommended calibration period has been
extended. The chemistry must have a valid current calibration to be extended.
Within-Lot Pending
Indicates calibration has been requested for another cartridge with the same lot
number. Applies only to chemistries with within-lot capability.
Disable
System-initiated or operator-initiated; allows analysis to proceed on all other
chemistries without having to deprogram the disabled chemistry. To continue to
process a disabled chemistry, the operator must enable the chemistry. When the
disabled status is removed, the chemistry returns to the current calibration status.
n/a
Indicates calibration does not apply.
Table 4.2
Calibration Status Messages
(Continued)
Message
Explanation
Table 4.3
Conditions and Highlights for Calibration Status Warnings
IF the Calibration Status has...
THEN the Highlight
is...
Which Means...
Cal Required
Red
The chemistry cannot be processed
now.
Chem Bypassed
Cal Failed
Cal Timed Out
Disabled
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......