Reagent Load/Calibration
Calibration Status
1
Select
Rgts/Cal
from the menu bar.
2
Select
Assign F7
.
3
Use the down arrow option button to open list of calibrators.
4
Select the calibrator.
5
Enter the ID of the calibrator or the rack/position number.
•
If using a bar code label, the label must match the assigned calibrator ID.
•
If a rack/position is assigned, the calibrator must be placed in this assigned position.
6
Repeat Steps 3–5 to assign additional calibrators.
7
When calibration assignment is complete, select
Close
.
Calibration Status
Introduction
Calibration status, as well as reagent status, is accessed from the
Rgts/Cal
dialog box. Calibration
status is shown to determine which chemistries require calibration.
Calibration Status Messages
Table 4.2
Calibration Status Messages
Message
Explanation
Requested
Indicates that calibration is pending and calibrators have been identified.
Calibrated
Indicates that the chemistry has been calibrated and has not exceeded its
calibration time period.
Cal Required
Appears when:
•
a new cartridge is loaded.
•
a user-defined cartridge is loaded.
•
a non-overrideable calibration failure has occurred.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......