Reagent Load/Calibration
Reagent Load
4
•
can be loaded using the hand-held bar code reader or loaded manually.
Load Reagent Parameters
1
Select
Rgts/Cal
from the menu bar.
NOTE
If parameters are loaded as part of the reagent cartridge load, proceed directly to Step 3.
2
Select
Params F2
.
NOTE
Parameters cannot be loaded if the reagent cartridge is onboard and the system status is
Running
.
3
After the reagent cartridge loads successfully, a
Load Reagent Parameters
dialog box appears.
Enter data for cartridges that require parameter entry.
To automatically load reagent parameters:
•
Use the hand-held bar code reader to scan each of the four bar codes.
•
The blue bar indicator on the number button indicates the parameter was scanned
successfully.
Unacceptable bar codes are not read and a message appears.
4
To manually load reagent parameters:
•
Type the parameter information, from the parameter card, on each line of the
Load Reagent
Parameters
dialog box.
NOTE
Readable parameters coincide with each bar code on the card.
•
The blue bar indicator on the number button indicates the parameter was entered
successfully.
5
Select
Done
to exit and save information.
OR
Select
Clear
to delete information.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......