A13914AF
4-3
Reagent Load/Calibration
Reagent Load
4
Sort Reagent Status
The order in which CC reagents are shown on the
Reagent Status/Calibration Status
screen can vary
according to operator selection.
1
Select
Rgts/Cal
from the menu bar.
2
Press
(
Page Up
)
or
(
Page Down
)
to go to pages 3–8.
3
Select
Sort F9
. A dialog box appears.
4
Select the sorting option in which the CC Chems should be shown.
•
Position (on reagent carousel).
•
Chemistry (Lists CC Chems alphabetically. The oldest cartridge is listed on top, for multiple
carts of the same Chem.)
•
Days Left (onboard stability or expiration date, whichever comes first).
•
Tests Left (number of tests left in cartridge).
•
Cal Time Left (oldest Cal date listed first).
NOTE
The option selected is shown in ascending or alphabetical order. When the sorting option is
selected, the sorting order is retained until the operator changes it.
Level Pending
Reagent cartridge has been loaded and is waiting to be level-sensed.
Level Sense Error
Level sense check failed to properly detect reagent in one or more
compartments.
n/a
Reagent status is not applicable for this chemistry (for example, an
electrolyte chemistry).
Table 4.1
Reagent Status Messages
(Continued)
Message
Explanation
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......