Operation Principles
Parameters
Figure 3.10
Time vs Fluorescence plot
The minimum stop time is 10 seconds, the maximum stop time (maximum duration) is
1,200 seconds and the default stop time is 300 seconds (5 minutes).
When you assign the TIME parameter to a plot axis, the divisions on the axis change accordingly.
To find the time (in seconds) per channel in a one-parameter histogram, divide the stop time (in
seconds) by 1,024 (0.001 second = 1.0 ms).
For a two-parameter plot, divide by 64, 128, 256 or 512 depending upon the plot resolution you are
using.
RATIO Parameter
The RATIO parameter is calculated, not acquired directly. When you select a parameter, you specify
which signal is the numerator and which is the denominator.
A ratio of 1 is at channel 1,023. If you assign RATIO to a plot axis, RATIO events appear at a lower
channel if the intensity of the numerator signal is less than the denominator signal.
To calculate the actual ratio at a particular intensity for a one-parameter histogram, divide the
intensity by 1,024. For a two-parameter plot, divide by 64, 128, 256 or 512 depending upon the plot
resolution you are using.
RATIO
Numerator
Denominator
------------------------------------
1024
×
=
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......