Operation Principles
Protocols
Signals Generated
The signals available are dependant upon whether you have a 6, 8, or 10 color configuration. INT
stands for Integral and TOF stands for Time-Of-Flight.
•
FS INT, FS PEAK, FS TOF
•
SS INT, SS PEAK, SS TOF
•
FL1 INT, FL1 PEAK, FL1 TOF
•
FL2 INT, FL2 PEAK, FL2 TOF
•
FL3 INT, FL3 PEAK, FL3 TOF
•
FL4 INT, FL4 PEAK, FL4 TOF
•
FL5 INT, FL5 PEAK, FL5 TOF
•
FL6 INT, FL6 PEAK, FL6 TOF
•
FL7 INT, FL7 PEAK, FL7 TOF
•
FL8 INT, FL8 PEAK, FL8 TOF
•
FL9 INT, FL9 PEAK, FL9 TOF
•
FL10 INT, FL10 PEAK, FL10 TOF
The software allows you to select a maximum of 16 parameters, including derived parameters. Log
and linear scaling of the Integral and Peak signals are also available but when Lin and Log
parameters are selected for the same signal (Peak or Integral), this will be counted as one parameter.
TOF is available in linear only.
Protocols
A Protocol is a collection of information about how the Cytometer and Navios software is set up.
Protocols can be created for acquiring samples or analyzing stored data. It contains the following
information:
•
Display configuration
•
Parameter names and configuration
•
Regions
•
Gates
•
Color definitions
•
Statistics
•
Instrument settings
•
Reporting Templates (FlowPAGEs)
Special Protocols and Panels
QC Protocols (Export): Any Protocol name beginning with the characters “QC” (upper-case only)
over-rides the current Export Data Format setting on the Workspace Preferences / Publish tab
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......