Replace/Adjust Procedures
Replace Reagents
Replace Reagents
About the Reagent Containers
Reagent Container Capacity
The internal sheath fluid container has a working capacity of about 500 mL. It is automatically
replenished from the external 10 L sheath fluid container. When you fill a completely empty sheath
fluid container (after cleaning or replacement), you need about 1 L of sheath fluid due to
pressurization and level sensing requirements.
NOTE
A bottle of IsoFlow sheath fluid holds 1.8 L.
Cleaning Agent Container
The cleaning agent container has a working capacity of about 500 mL. This is the amount of reagent
needed when you are filling the cleaning agent container after
Cleanse Level Warning or
Cleanse
Level Error appears. When you fill a completely empty cleaning agent container (after cleaning or
replacement), you need about 1 L of cleaning agent due to pressurization and level sensing
requirements.
NOTE
bottle of FlowClean cleaning agent holds 500 mL.
Replace The 10 L External Sheath Fluid Container
Perform this procedure whenever:
•
The
Sheath Low
indicator is red.
•
The Cytometer has an external 10 L
sheath fluid container and internal
containers for cleaning agent (1) and
sheath fluid (2).
•
For best use of reagents, refill the internal
reagent containers only when the
instrument indicates that they are low.
•
If you replace a reagent container, clean it
before you put it into the instrument and
fill it. See
Clean The Internal Sheath Fluid
C
B
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......