Cleaning Procedures
Clean The Sampling System
7
Carefully push the moistened gauze pad up against the inside of the MCL sample head
(1 )
and
scrub away any debris inside and around the sample probe.
8
Continue scrubbing the sample head and probe by pushing the head up and down 10 times
during a 60-second period. Replace moistened gauze as needed.
9
Rinse the MCL sample head and probe with gauze moistened with water.
10
Moisten a Q-tip with distilled water.
b