Using Navios Software
Creating And Importing Panel Report Templates
10
Verifying A Panel Report Template
Before using the report template, verify that it is working as expected. Check that the result lines
are organized in the desired sequence.
NOTE
The row that an equation is created on must be constructed using information from rows above the
equation row.
NOTE
You must save a protocol
prior
to using it as part of a panel.
To verify a panel report template after creating or editing it:
1.
Print a copy of the panel report template.
2.
Verify that the report includes all desired lines in the proper sequence.
3.
Verify that the equations are referencing the appropriate rows and that these rows are
above
the equation row in the template.
Run a test panel on a normal sample and verify each result printed on the panel report. If there is
an inconsistency, verify the report template again for proper formulas and references; then verify
again.
Global Tab Settings
1
Allows you to define a global report comment that prints on
every
panel report generated from this
template.
NOTE
You can edit the comment on individual reports if necessary. See
2
Allows you to select the reporting unit format: SI Units or US format (which is the default if SI Units is
not selected).
3
Allows you to choose whether or not to print ranges on the report.
4
Allows you to choose Cells/
μ
l CAL, Cells/
μ
l Hem or Number. Cells/
μ
l CAL adjusts absolute counts based
on the CAL factor. Cells/
μ
l Hem calculates absolute counts based on the Cell Population selected from
the hematology patient information and the number of events in the result region. Number displays
the unadjusted count in the region.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......