PN A51387A
B
3-20
Installing and Calibrating an Aperture Tube
Verifying the Calibration
5.
When the run is complete, the Verify Aperture Calibration window opens.
6.
In the Verify Aperture Calibration window:
•
If desired, click
(
Find Peak
)
to place cursors at or on either side of the peak of the
calibration graph. If you are performing a modal calibration, clicking
(
Find Peak
)
automatically places a cursor on the modal channel. If you are performing a mean
calibration,
(
Find Peak
)
automatically places two cursors on the graph, one on either side
of the peak. The cursor(s) define the mode or median used to calculate the new Kd.
•
If desired, click
(
)
to print the window. This button does not display a dialog box or
allow you to change print settings.
Table 3.5
Verifying Aperture Calibration
If the window displays...
Then...
The Calibration is OK
The measured particle size is within 3% of the actual
calibrator size, and the calibration was successful.
Click
(
Close
)
to exit the window.
The instrument needs to be
recalibrated
The measured variation is too great, and the calibration
was not successful.
1.
Click
(
Close
)
to exit the window.
2.
In the Instrument Toolbar, click
(
Reset
)
.
3.
If necessary, adjust the amount of calibrator in the
electrolyte solution.
4.
Restart the calibration procedure.
Summary of Contents for Multisizer 4
Page 12: ...xii List of Figures ...
Page 14: ...xiv List of Tables ...
Page 22: ...PN A51387AB xxii Introduction Before You Start Warnings and Cautions ...
Page 42: ...PN A51387AB 1 20 Analyzer Overview Preparing the Analyzer for Sample Runs ...
Page 160: ...PN A51387AB 6 16 Analyzing a Sample Using Blank Runs ...
Page 232: ...PN A51387AB 10 8 Regulatory Compliance Additional Security Features ...