Installing and Calibrating an Aperture Tube
Calibrating the Aperture
3
Completing the Calibration Run (Accepting the Kd)
To complete the calibration run:
1.
When the calibration run is complete or stopped, a Calibrate Aperture window opens.
2.
In the Calibrate Aperture window, the Kd pane displays the Old and New Kd values.
Table 3.3
Pausing / Stopping Calibration
Click...
To...
(
Pause
)
Pause the calibration.
(
Stop
)
Stop the calibration. The Multisizer 4 software
displays calibration results from the stopped run in
the Calibrate Aperture window.
(
Cancel
)
Cancel the calibration. The Multisizer 4 software does
not display calibration results. To restart calibration,
select
Calibration
>
Calibrate Aperture
on the Main
Menu bar.