Installing and Calibrating an Aperture Tube
Using Small Aperture Tubes (20 µm and 30 µm)
If you are unable to obtain results similar to those in
after thoroughly cleaning the system
(see
, page
), your aperture tube may be damaged or blocked.
NOTE
is an example of a clean noise profile. Factors such as aperture size, current and gain
settings, and electrolyte can affect results. The noise profile graphs below show the relative
difference between a system that is performing well
and one that may need cleaning
For information on subtracting a blank run to obtain a noise profile, see
, page
.
Figure 3.4
Noise Profile, Clean System (20 µm aperture)