PN B05577AC
6-76
Sample Modules
Universal Liquid Module
Loading an SOP
To load an SOP:
1
Run
.
2
Select
Load an SOP
. SOPs can also be loaded from the status bar by clicking on the
Load SOP
button.
3
Select the desired SOP file from the SOP subfolder or the subfolder where the file was last saved.
Suspension Fluids
The materials used in the construction of the ULM that come in contact with the suspension fluid
are Teflon
®
, 316 stainless steel, glass, and Kal-rez
®
.
Only use suspension fluids that have flash points greater than 10.0°C (50°F), with specific gravities
greater than or equal to 0.6, and that meet the requirements listed below. Examples of fluids that
can be used in the ULM are water, acetone, weak aqueous-based acids or bases, mineral oils,
ethylene glycol, hexane, etc. The suspension fluid used depends on the type of sample to be
analyzed.
Suspension fluid requirements:
•
Chemical Compatibility
- does not adversely react with the ULM components mentioned above.
•
Transparency
- transparent in the 450 nm to 900 nm range
•
Clarity
- no contamination or extraneous particles
•
Consistency
- no bubbles
•
Density
- lighter than the sample particles
•
Wettability
- able to deflocculate the sample particles
•
Fluidity
- low viscosity for rapid stirring
•
Insolubility
- no dissolution of particles
•
Chemical neutrality
- neutral to the sample and the ULM
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......