PN B05577AC
6-61
Sample Modules
Tornado Dry Powder System
6
Figure 6.38
Sample Cell Windows
2
Fold three to four lens tissues in half.
3
Apply lens cleaner onto the creased area of the folded lens tissue.
4
Place the lens tissues on the window surface, holding the tissue in place with two fingers.
5
Slowly, wipe the window.
6
Discard the used lens tissue.
7
Re-assemble the windows and sample cell.
Cleaning the Sample Tubing
The lines that carry the sample into the sample cell may need to be cleaned separately when the
sample cleaner cannot clean them completely.
1.
Screws
2.
Sample Cell Windows
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......