MLM Troubleshooting
Tornado Dry Powder System
Problem
Possible Cause/Solution
No communication with optical bench
•
Module docked incorrectly. Eject module and re-dock.
Stirrer will not turn
•
Liquid too viscous
•
Cell not mounted properly
•
O-rings missing from stirrer bar
Sensor light not on
•
Module not docked properly. Eject module and re-dock.
•
Cell not mounted properly
Obscuration too high
•
Dirty windows. Clean windows.
Module leaking
•
Check fittings
The Tornado Dry Powder System is intended for use with the
LS 13 320 Optical Bench. It is capable of feeding and measuring
dry powder samples in the size range 0.4 µm to 2000 µm. The
Tornado DPS measures the entire sample presented to the
instrument, with a sample volume that allows an analysis that
will provide statistically accurate results. The range of sample
volumes the instrument can accept should provide for a
minimum 10 second run at controlled obscuration. There is no
need for sample preparation and operator intervention is
minimal. The system is controlled via software commands that
make it completely hands-free. The sample is placed in a sample
holder and delivered to the sensing zone in the optical bench by
a vacuum.
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......