PN B05577AC
6-31
Sample Modules
Micro Liquid Module
6
Figure 6.15
Micro Liquid Module
The MLM measures the entire sample in the instrument by circulating the sample with a stir bar.
The amount of sample needed depends on its size and concentration. Prior sample preparation is
often needed to achieve proper dispersion (see
for more information
on how to disperse samples) in order to obtain accurate and valid results.
MLM Description
The Micro Liquid Module is a chemically resistant, integrated system consisting of:
•
A sample vessel (Micro Liquid Cell) with stirrer to introduce the sample to the system
•
Main housing consisting of stirrer control and magnetic stirrer motor
The system is designed to work with small quantities of suspension fluid (12 ml), both aqueous and
non-aqueous.
Figure 6.16
Micro Liquid Module
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......