PN B05577AC
6-12
Sample Modules
Aqueous Liquid Module
7
Step 5 provides the user with a summary of the options selected in the previous steps. The SOM
is saved during this step.
IMPORTANT
SOM files, by default, are saved under the SOP folder, but they can be saved to the folder of
your choice as well.
Under high security (21 CFR Part 11, for compliant version of software only), SOMs can only be saved
by the Administrator. Under Operator mode, the
Save
button in step 5 of the SOM creation is not
displayed.
8.1.3.3 Loading an SOM
To load an SOM:
1
Run
.
2
Select
Load an SOM
.
3
Select an SOM file from the list.
When an SOM is loaded, the status panel will display information about the loaded SOM. The
Start
Cycle
button will change to
Start SOM
. Clicking on this button will start an analysis using all the
settings unique to that SOM. Clicking on the
Remove SOM
button will bring back the
Start Cycle
option.
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......