PN B05577AC
5-7
Data File Menus
Edit Menu
5
Edit Menu
Figure 5.3
No Security
Set Up With Security
A setup with security selection for the Edit menu would have “Add Electronic Signature” displayed.
Copy Window
This command will place the information displayed in the LS 13 320 window onto the clipboard.
Edit Sample Info
Information about the sample can be edited through this option. Under all levels of security, except
No Security, any changes made during the editing process will be recorded in the File History.
NOTE
Expect that all edit options change depending on the level of security.
Under No Security any user can edit the data file. For Low Security the operator and reviewer may
edit data only if the administrator selects the options under the user privileges dialog. Medium
Security does not allow either the operator or reviewer to edit data. For High Security see Security
Options.
Summary of Contents for LS 13 320
Page 4: ...PN B05577AC iv Revision History ...
Page 12: ...PN B05577AC xii Safety Notice Scope of Manual ...
Page 30: ...PN B05577AC xxx LS 13 320 Introduction System Components ...
Page 57: ...PN B05577AC 2 21 Installation Making Measurements 2 Figure 2 18 Run Cycle Options Dialog ...
Page 59: ...PN B05577AC 3 1 CHAPTER 3 LS 13 320 Software Figure 3 1 LS 13 320 Software ...
Page 80: ...PN B05577AC 3 22 LS 13 320 Software Preference Options ...
Page 86: ...PN B05577AC 3 28 LS 13 320 Software Preference Options ...
Page 116: ...PN B05577AC 4 10 Regulatory Compliance 21 CFR Part 11 Starting Security Enabled Software ...
Page 169: ...PN B05577AC 6 39 Sample Modules Micro Liquid Module 6 Figure 6 27 SOM List Dialog ...
Page 216: ...PN B05577AC 6 86 Sample Modules Universal Liquid Module ...
Page 228: ...PN B05577AC B 10 Sample Handling Diluent Selection ...
Page 238: ...PN B05577AC C 10 Optical Models Statistics ...
Page 252: ...PN B05577AC Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 253: ......