Instrument Specifications
Temperature and Humidity
IMMAGE 800 Operations Manual A11403
System Description
September 2016
Page 2-47
2
Temperature and Humidity
Drain Requirements
Flow Rate: 3 Liters/hour minimum
Waste Container Placement: The opening should be no higher than the top of the
instrument.
Regulatory Agency Approvals
The IMMAGE 800 meets the safety requirements for the following agencies: CE, UL,
CSA, IEC and CENFLEC.
Environmental Conditions
System can operate up to 8000 ft. (2,438m) elevation.
Capacities
The following table lists various system capacities.
Maximum Leakage Current
120V, 60 Hz: 40 uA
240V, 50 Hz: 68 uA
WARNING
IF THE EQUIPMENT IS USED IN A MANNER NOT SPECIFIED BY BECKMAN COULTER,
INC., THE PROTECTION PROVIDED BY THE EQUIPMENT MAY BE IMPAIRED.
Ambient Temperature
+15°C to +32°C
Ambient Relative Humidity (RH)
15% to 85% (non-condensing)
Reagent Compartment Temperature
13-22°C (32°C Ambient, <45%RH)
Reaction Module Temperature
+37°C ± 0.5°C
Table 2.9
System Capacities
Item
Capacity
Reagents
24 reagent cartridges can be loaded.
Reagent cartridge
40, 150, or 300 tests per cartridge.
Reaction buffers
4 bottles can be loaded.
Buffer bottle
120 mL: 350 tests.
Samples
72 samples can be loaded.
Sample diluents
4 bottles can be loaded.
Diluent bottles
120 mL: number of dilutions is workload dependent.
Sample dilution segments
4 segments of 36 wells each.
Dilution well
300 µL.