Overview
Operating in Host Query Mode
Sample Programming
IMMAGE 800 Operations Manual A11403
Page 7-80
September 2016
3
NOTICE
If a position is programmed by a bar coded tube with host query, the
position will automatically clear when a new bar coded tube is read in
that position. The new bar coded tube can then run as a result of the
host query for that position.
All manually programmed positions must be cleared manually in order
to run any other samples in those positions. If a bar coded tube with a
host query is run in a position previously manually programmed, the
position will not automatically clear. A conflict warning message will
display.
If samples are...
then...
bar coded
load samples in any rack and position.
not bar coded
program rack and position, and sample ID for
each sample. Load samples in the
programmed racks and positions. (Refer to
Programming a Sample in this chapter.)
(Refer to CHAPTER 2,
System Description
4
Select
Main
from the menu bar.
AND
Select
Run
.
5
Select
<OK>
in the Check Dilution Segments dialog box to start the run.
OR
Select
<Cancel>
to exit without starting the run.
Step
Action
, continued
(2 of 2)