Requesting and Canceling Calibration
Assigning Calibrator Rack and Position
IMMAGE 800 Operations Manual A11403
Reagents/Calibration
September 2016
Page 6-29
6
Figure 6.11
Clear Racks Dialog Box
Assigning Calibrator Rack and Position
A non-bar coded calibrator must be assigned a rack and position when requesting
calibration.
Follow the steps below to assign rack and position to a non-bar coded calibrator
sample.
Canceling a Requested Calibration
Follow the steps below to cancel a requested calibration.
E010237S.EPS
Step
Action
1
From the Request Calibration screen type the
Available
rack and position
for the calibrator in the
Rack
and
Position
fields.
2
Select the options button
<
▼
>
beside the desired position for the
calibrator.
3
From the Calibrator Lots dialog box, select the appropriate calibrator lot
number.
OR
Select
<Cancel>
to return to the Request Calibration screen without
selecting a calibrator lot number.
4
Select
Save [F9]
to save the calibration request and assignment.
OR
Select
Cancel [F10]
to return to the Reagent Status/Calibration Status
screen without saving.
Step
Action
1
Select
Rgts/Cal
from the menu bar.
2
Select the button(s) to the left of the requested chemistries to be canceled.
3
Select
Cancel Request [F7]
.
The Cal Status of the canceled chemistries will return to the status prior to
calibration request.