Printing UDR Reports
Description of UDR Reports
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-112
September 2016
Printing UDR Reports
Description of UDR Reports
The following table describes the UDR report options. Refer to APPENDIX B,
for examples of the UDR Reports.
Table 5.16
UDR Reports
Report
Description
Definition Report
Identifies UDR chemistry definition protocol and
calibration information.
Calibration Results Report
Identifies UDR instrument response results per
replicate of each UDR calibrator for the selected
chemistry.
Calibration Report
Identifies UDR mean instrument responses per
replicate of each UDR calibrator for the selected
chemistry, the curve-fit model, and associated
parameters.
A calibration report can be printed once a UDR
chemistry is defined and before a calibration is
requested. The printed report will be blank.
Single-Point Update
Calibration Results Report
Identifies UDR instrument response results per
replicate of single-point update calibrator for the
selected chemistry.
Single-Point Update
Calibration Report
Identifies UDR mean instrument responses per
replicate of single-point update calibrator for the
selected chemistry, the curve-fit model, associated
parameters, and scale factor.
Plot Robust Means Data
Report
Identifies UDR robust means data for a single
replicate of a selected UDR calibrator or rate mode
report.
Rate Mode Results
Identifies UDR instrument responses per replicate
of each UDR sample for the selected chemistry.
Patient Results
Identifies UDR concentrations per replicate of
patient sample programmed.