Diagnostics
368
CEQ™ 8000 Genetic Analysis System
Monitoring the Baseline
You must perform the optical alignment procedures prior to monitor-
ing the baseline (“Performing an Optical Alignment’ on page 333).
To monitor the baseline, perform the following steps.
1. Select
Run | Monitor Baseline
from the Run Module main menu.
2. From the
Monitor Baseline
dialog box:
a. Select the
Enable Monitor Baseline
check box.
b. Select the
Autosave
check box (to save monitor baseline data).
c. Enter a name in the
Name
field.
d. Select a
Project
from the drop-down menu and then select
OK
.
5. To view the baseline trace:
a. Access the desired Analysis module.
b. Select
File | Open
from the menu.
c. Click on the
Sample Data
tab and select the desired baseline data (the
baseline data is displayed).
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...