Direct Control Procedures
User’s Guide
331
Figure 195:Loading the Sample Plate
3. When finished positioning the plate, close the Sample Access Cover and then click
the
Load
button of the
Capillaries Exposed
The separation gel within the capillaries will dry out if the capillaries
are left exposed to the air for more than 15 minutes.
Figure 196:Capillaries Exposed Dialog
Loading the Buffer Plate and Evaporation Cover
1. Align the notched corner of the Buffer Plate with the alignment line on the Buffer
2. Gently push the Buffer Plate towards the front of the instrument and then set the
plate into the transport.
900500L AI
900497e.AI
Sample Pate
Holder
Sample Plate
Guide Pin
Sample Plate
Sample
Plate
Notched Edge
align
900497e.A1
900500e.AI
Wetting Tray Retainers
CAUTION
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...