134
CEQ™ 8000 Genetic Analysis System
Condition Method
The
Condition
method (see Figure 78) should be used before a run if the system has
not been used for a long period of time. Performing this method will fill the capillaries
with gel 6 times. Then, after separation, thoroughly purge the manifold with fresh gel
followed by a capillary fill.
Thoroughly clean the wetting tray with D.I. water before every run to remove
old or dried gel.
Creating or Editing a Method
To create and/or edit a method, perform the following steps.
1. Select
Edit | Method
from the Sample Setup Main menu.
2. Highlight the desired method in the
Choose Method to Edit
dialog box and
click
OK
.
3. In the
Method - Capillary Temperature
dialog box:
a. Enter a temperature between 35 and 60° C.
b. Select the
Wait for Cap Temp
option (if desired).
c. Click
OK
to exit the
Method
dialog box or continue with step 4 to make
additional changes to the method.
4. Select
Denature
from the
Event
list and then:
a. Enter a duration between 0 and 180 seconds (if 0 is selected, there will be no
denaturation).
b. Click
OK
to exit the
Method
dialog box or continue with step 5 to make
additional changes to the method.
3. Select
Pause
from the
Event
list and then:
a. Enter a time duration between 0 and 10 minutes.
b. Click
OK
to exit the
Method
dialog box or continue with step 6 to make
additional changes to the method.
3. Select
Inject
from the
Event
list and then:
a. Enter an injection voltage between 0.1 and 12.0 kV.
b. Enter a time duration in seconds.
c. Click
OK
to exit the
Method
dialog box or continue with step 7 to make
additional changes to the method.
4. Select
Separate
from the
Event
list, and then:
a. Enter the separation voltage between 0.1 and 12.0 kV.
b. Enter a time duration in minutes.
c. Click
Advanced
to specify the two separation phases for the method. Click
OK
to return to the
Method
dialog box.
Summary of Contents for CEQ 8000
Page 42: ...Program Description 28 CEQ 8000 Genetic Analysis System...
Page 98: ...84 CEQ 8000 Genetic Analysis System...
Page 110: ...96 CEQ 8000 Genetic Analysis System...
Page 120: ...106 CEQ 8000 Genetic Analysis System...
Page 128: ...114 CEQ 8000 Genetic Analysis System...
Page 152: ...138 CEQ 8000 Genetic Analysis System Figure 80 Report Format dialog...
Page 154: ...140 CEQ 8000 Genetic Analysis System...
Page 162: ...Run Procedures 148 CEQ 8000 Genetic Analysis System...
Page 220: ...Sequence Analysis Procedures 206 CEQ 8000 Genetic Analysis System...
Page 318: ...Fragment Analysis Procedures 304 CEQ 8000 Genetic Analysis System...
Page 329: ...Exporting Database Items User s Guide 315 Figure 180...
Page 364: ...Direct Control and Replenishment 350 CEQ 8000 Genetic Analysis System...
Page 380: ...Routine Maintenance 366 CEQ 8000 Genetic Analysis System...