![Beckman Coulter Biomek NXP Preinstallation Manual Download Page 21](http://html1.mh-extra.com/html/beckman-coulter/biomek-nxp/biomek-nxp_preinstallation-manual_2719713021.webp)
B20063AC
1-7
Preparing the Site
Air, Vacuum, and System Fluid Requirements
1
Beckman Coulter’s compressor option, PN C25555, can be ordered in conjunction with a booster
pump bypass fittings kit. The compressor requires a 220V 50/60 Hz single phase power supply. It
should be noted that air compressors are louder during operation than shop air.
Figure 1.1
Recommendation for an Air Pressure Regulation System Configuration
Vacuum Supply Requirements
If conducting PCR or Nucleic Acid Preparation applications that rely on vacuum filtration, the site
must have the following vacuum supply available:
•
For the SPE filtration manifold, a minimum 20 in. Hg @ 4.5 SCFM (67.7 kPa @ 0.127 m3/min) is
required.
1.
Compressed Air
2.
Quick Disconnect Fitting (supplied)
3.
1/4" OD Tubing (supplied)
4.
Biomek NX
P
Instrument
5.
1/4" Input
6.
Filter/Regulator