2-16
Operation
Setting Up a Zonal Run
Steps 1 and 2 are saved under the selected program number.
7
Press the
EXIT
soft key to exit program mode.
Running a Program
1
Press
OPT
.
The
Prog Recall:
field flashes.
2
Use the keypad to enter the program number in the
Prog Recall:
field.
3
Press
ENTER
, then
START.
Setting Up a Zonal Run
When the centrifuge is in the zonal mode (key switch is in the
ZONAL
position), open-door operation
at speeds up to 3000 RPM is enabled to allow loading and unloading of the rotor while it is spinning.
The JCF-Z continuous flow/zonal rotor is the only rotor that can be used while the key switch is in the
ZONAL position
. See the JCF-Z rotor manual for complete rotor operating instructions.
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Accel : 5 Decel : 5
Prog Recall: Setup: 1 Step #1
STEP1
STEP2
SAVE
EXIT
Rotor : JA-30.50
Speed : 26 000 RPM = 81 800 x G
Time : 00:20
Temp : 20°C Max Temp : 24°C Shutdown? No
Switch steps, save, or exit program mode
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Accel : 5 Decel : 5
Prog Recall: 1 Setup: Step #1
STEP1
STEP2
EXIT
Rotor : JA-30.50
Speed : 26 000 RPM = 81 800 x G
Time : 00:20
Temp : 20°C Max Temp : 24°C Shutdown? No
Enter program number (1Ð30) .
Summary of Contents for Avanti J-26S XPI
Page 4: ...PN B10093AE iv Revision History ...
Page 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Page 14: ...xiv Illustrations ...
Page 16: ...xvi Tables ...
Page 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Page 36: ...PN B10093AE 1 16 Description Available Rotors ...
Page 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Page 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Page 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 87: ......